LBGF-8000A/B High-Precision Fully Automatic Industrial Analyzer
Other Products
Model LBGF-8000A: Tests three indicators—moisture, ash, and volatile matter—in a single furnace.
Model LBGF-8000B: Tests the volatile matter indicator, designed according to the national standard method for volatile matter testing.
When LBGF-8000A and LBGF-8000B operate simultaneously, testing speed is faster and accuracy is higher, with the option to use either the rapid method or the national standard classic method.
Features of the High-Precision Fully Automatic Industrial Analyzer:
■ High Automation: After placing the sample, simply click the mouse to automatically complete the test and print results, making operation simple.
■ Accurate Testing: Tests on over twenty standard coal samples have yielded accurate results; comparative tests with the national standard method confirm that testing precision meets requirements. The LBGF-8000A and LBGF-8000B instruments adopt a dual-structure design, with volatile matter testing as a separate part of the instrument, while the other part tests moisture and ash. Test results exceed national standard requirements.
■ High Efficiency:
Rapid Method: Moisture, ash, and volatile matter for 19 samples ≤120min (LBGF-8000A/B)
≤180min (LBGF-8000A)
Classic Method: Moisture (dryness inspection) and ash (slow ash + dryness inspection) for 19 samples: 260min; volatile matter: 120min (LBGF-8000A/B).
■ The three indicators—moisture, ash, and volatile matter—can be measured in any combination or individually.
■ Utilizes imported integrated furnaces for fast heating, energy savings, and short testing times.
■ Employs imported electronic balances and stepper motors to ensure stable testing performance.
■ New feature: Pause experiments midway and resume the next day. New electric push rod for automatic lid opening and closing. New wireless data sharing function allows users to wirelessly transmit data directionally.
■ Can connect to electronic balances.
Technical Parameters of the High-Precision Fully Automatic Industrial Analyzer:
Power:
LBGF-8000A ≤5KW
LBGF-8000B ≤4KW
LBGF-8000A/B ≤9KW
Gas Requirements:
Oxygen: Purity 99.5%, pressure after减压 0.1Mpa
Nitrogen: Purity 99.5%, pressure after减压 0.1Mpa
Pressure Reducer: 0-25Mpa
Low End 0-0.4Mpa
Crucible Quantity: 20, can test 19 samples at once
Sample Mass: Moisture and ash samples 0.6-1.1g (adjustable); volatile matter 0.9-1.1g
Furnace Temperature Range: Room temperature to 0-1000℃
Analysis Precision: Meets the requirements of GB/T212-2008 "Proximate Analysis of Coal" standard.
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 鹤壁蓝博 |
| Spec: | |
| Stock: | |
| Origin: | China / Henan / Hebishi |